(4) Detailed Sequence of Driven Shaft Grip Release (Fast/HR)
Fig. 5.9.4 Driven Shaft Grip Release (HR/ST)
Preprocessing: Fig. 5.9.1
START
Driven shaft grip release driving
DRIVE (MZ3/ON)
TZ3 START
Timer START
Is SZ3 opened within TZ31?
N
Y
Driven shaft grip release stop
DRIVE (MZ3/OFF)
(hard)
Timer STOP
Timer > TZ33 or Timer <TZ32
Y
N
END
(5) Detailed Sequence of Driven Shaft Grip (HR/ST)
Fig. 5.9.5 Driven Shaft Grip (Fast/HR)
Preprocessing: Fig. 5.9.1
START
Driven shaft grip driving
DRIVE (MZ3/ON)
Is SZ3 closed within TZ34?
N
Y
Driven shaft grip stop
DRIVE (MZ3/OFF)
(hard)
END
NOTE: Slow is used only for self-diagnostics.
009-051-04
05.31.99 FM2460
3
4
5
6
Driven shaft grip release stop
DRIVE (MZ3/OFF)
(soft)
Level 0 error
Driven shaft grip release
03C7
error
Error log (detailed information)
Timer data recording
Level 2 error
Driven shaft grip release
23C7
operation error
Driven shaft grip stop
DRIVE (MZ3/OFF)
(soft)
Level 0 error
03C9
Driven shaft grip error
CR-IR341 Service Manual
7
8
9
10
Table 5.9.4 Operation Parameters
Symbol
Driven shaft grip timeout
TZ31
Lower limit for driven shaft grip release spec (ST/HR)
TZ32
Upper limit for driven shaft grip release spec (ST/HR)
TZ33
FR1H0433.EPS
Table 5.9.5 Operation Parameters
Symbol
Driven shaft grip timeout (fast/HR)
TZ34
FR1H0434.EPS
MD - 147
Description
(ST/HR)
Value
Description
2 / 10 sec
TR1H0084.EPS
Value
5 / 10sec
1.2 /3.0sec
2.4 /4.9sec
TR1H0083.EPS
MD - 147