FujiFilm CR-IR341 Service Manual page 680

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
Replacement
(1) Remove the following covers.
Lower front cover, lower left-hand side cover, lower left-hand side transparent cover
"3. Removing and Reinstalling Covers"
(2) Remove the Kapton
"10.12 Checking and Replacing Kapton
(3) Disconnect the connectors (CN1-3) of MZ1.
(4) Remove MZ1.
(3)
CN1, 2, 3
(5) Install a new MZ1 by reversing the removal steps.
NOTE
When installing MZ1, the same settings for S1-S6 of a new MZ1 should be made as
those for the older one.
(6) Perform the procedures described in "Check after MZ1 Replacement."
Check after MZ1 Replacement
(1) Power ON the machine.
Make sure that the machine starts up normally.
(2) Check MZ1 again to ensure that the motor operates normally.
(3) Check image performance.
"3.2 Checking Image" in Preventive Maintenance Volume
009-051-04
05.31.99 FM2460
3
4
5
6
®
belt.
(4)
MZ1
BR M4x8, x5
(2)
Spring
(1)
Left-hand side transparent cover
BR M4x8, x2
CR-IR341 Service Manual
7
8
9
10
®
Belt/Tensioner/Flywheel"
(2)
¤
Kapton
belt
Subscanning unit
FR1H1812.EPS
MC - 189
MC - 189