Remedy 2
Jam location:
Cause:
IP status:
Recovery procedure
a
Gain access from the front or right-hand side of the machine.
(1) Remove the upper front cover.
(2) Remove the tray of cassette set unit.
"4.1 Pulling Out and Pushing In Tray of Cassette Set Unit"
(3) Rotate the up-down IP removal unit rollers by hand to remove the IP.
b
Gain access from the front, right-hand, or left-hand side of the machine.
(1) Remove the upper front cover.
(2) Pull out the tray of cassette set unit.
"4.1 Pulling Out and Pushing In Tray of Cassette Set Unit"
(3) Remove the up-down conveyance guide of the up-down IP removal unit to take out
the IP.
009-051-04
05.31.99 FM2460
3
4
5
6
Up-down IP removal unit
Conveyance failure
The IP leading edge or trailing edge has not reached the before-side-
positioning conveyor.
a) The IP cannot be conveyed forward but can be conveyed backward.
b) The IP cannot be conveyed forward or backward.