Turning OFF the High-Voltage Switch (Software Switch)
"9. HV OFF" is displayed when the high-voltage switch is in the ON position, while "9. HV
ON" is displayed when it is in the OFF position.
(1) Touch the U-Utility button.
The screen switches to the U-Utility mode.
(2) Touch the upper left and upper right corners of the operation panel in sequence.
M-Utility is activated.
(3) Select "9. HV OFF."
The high-voltage switch is turned OFF, and "9. HV OFF" displayed on the screen
switches to "9. HV ON."
Display in
ON status
009-051-04
05.31.99 FM2460
3
4
5
6
M-Utility screen
0. QUIT
1. ERROR LOG UTILITY
2. CONFIGURATION SETTING
3. TEST MODE
4. ELECTRICAL UTILITY
5. SCANNER UTILITY
6. MECHANICAL UTILITY
7. FILE UTILITY
8. BACKUP MEMORY
9. HV OFF
10. MENU SETTING
>
CR-IR341 Service Manual
7
8
9
10
M-Utility screen
Display in
OFF status
0. QUIT
1. ERROR LOG UTILITY
2. CONFIGURATION SETTING
3. TEST MODE
4. ELECTRICAL UTILITY
5. SCANNER UTILITY
6. MECHANICAL UTILITY
7. FILE UTILITY
8. BACKUP MEMORY
9. HV ON
10. MENU SETTING
>
FR1H1416.EPS
MC - 5