FujiFilm CR-IR341 Service Manual page 262

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
3
4
ONTENTS
2.
Troubleshooting Procedures Based on Error Codes
2.1
How to Understand Error Code
A format of error code is shown below.
X Y Z Z
Hexadecimal notation
00-FF:
0-9, A-F: Error classification
0-3:
X: Error level
• FATAL error: 0
Error where the normal processing cannot be resumed.
It is necessary to troubleshoot and take remedial action immediately.
• WARNING: 1, 2, 3
Errors where the processing may be resumed by performing retry operation, etc.
This category includes an error that is merely logged as history information and an error
where the processing is resumed but its error code is displayed on screen.
Y: Error classification
0: OS (operating system software), CPU, library
1: Overall control, information gathering function,
output destination control function
2: Panel control
3: Conveyor-related control
4: Image processing related (reading)
5: Scanner control
6: Image processing related (printer)
7: IDT interface control
ZZ: Reference number
Managed according to the error classification
009-051-04
009-051-06
04.20.2001 FM3006
05.31.99 FM2460
5
6
Reference number according to error classification
Error level
8: Printer interface control
9: DMS interface control
A: ID information setup function
B: Network output image processing
C: FINP control
D: DICOM control
E: Undefined
F: Other (software install, version update, etc.)