03BC Side-Positioning Operation Error
Cause
MD1 is activated in order to detect the side-positioning home position (HP), but SD1
cannot detect the HP. (When SD1 is ON, it is detected as it transitions from ON to OFF
to ON. When SD1 is OFF, a transition from OFF to ON is detected.)
Supplementary explanation
During initialization, MD1 is driven to detect the HP of the latch assembly.
When SD1 is OFF during initialization, MD1 is driven in the HP direction, thereby
turning ON SD1.
When SD1 is ON during initialization, MD1 is once driven in the side-positioning
direction, thereby turning OFF SD1. MD1 is then driven in the HP direction again to
turn ON SD1.
As SD1 turns ON, the HP of the latch assembly is detected.
This error occurs when SD1 does not transition from ON to OFF, or from OFF to ON,
or from ON to OFF to ON, during the above operation.
Reference: If any trouble occurs on MD1/SD1 during side-positioning operation, "0422
Sensor related
Error-related I/O locations
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
009-051-04
05.31.99 FM2460
3
4
5
6
DSP Image Processing Error 1" occurs. After that, during initialization upon
power-ON again, 03BC error occurs.
If any trouble occurs on the MD1/latch assembly before side-positioning
operation during reading, "03C0 Reading Preparation Conveyance Error"
occurs. After that, during initialization upon power-ON again, 03BC error
occurs.
ON: Close, OFF: Open
SC2
SC1
SNS08A
CN
CN9
CN13
CN4
DRV03D
CN3
CN3
CN12
DRV08A
CN1
CR-IR341 Service Manual
Latch assembly
MD1 side-positioning motor
CNZ2
CND1
CNZ6
MD1
CNMD1
FR1H0035.EPS
MT - 96
SD1
side-positioning
Actuator
mechanism HP
for SD1
sensor
FR1H0034.EPS