1
2
C
ONTENTS
13E3 IP Grip Error
Cause
The IP that has been sucked and conveyed from the cassette is fed to the grip roller,
but SB2 does not turn ON.
Supplementary explanation
The IP that has been sucked and conveyed from the cassette is fed to the grip roller.
The IP fed to the grip roller is conveyed to SB2 by the driving of MB2.
As SB2 turns ON, the leading edge of the IP is detected.
This error occurs when the IP is fed to the grip roller but SB2 does not turn ON during
the above operation.
Sensor related
Error-related I/O locations
Cassette inlet
IP sensor
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
009-051-04
05.31.99 FM2460
3
4
5
6
ON: Close, OFF: Open
SB2
SNS08A
CN
CN13
CN4
DRV03D
CN3
CN3
DRV08A
CN1
CR-IR341 Service Manual
MB2
IP transport motor
FR1H0131.EPS
CNB1
CNB2
CNMB2
CNB3
MB2
FR1H0132.EPS
MT - 160
MT - 160