Replacement of Polygonal Mirror Assembly
CAUTIONS
• Use care not to let the polygonal mirror assembly fall or bump. If the polygonal mirror assem-
bly is shocked, its axis of rotation may be misaligned and thus it may become useless.
• Never touch the mirror face of the polygonal mirror (even when you are wearing gloves). If the
mirror face of the polygonal mirror is touched, the polygonal mirror will become unusable.
(1) Power OFF the machine.
(2) Pull out the scanner unit.
"7. Pulling Out and Pushing In Scanner Unit"
(3) Remove the scanning optics unit.
"9.3 Removing and Reinstalling Scanning Optics Unit,
(4) Remove the clamp.
(5) Cut the cable tie that retains the filter, and then remove the filter.
(6) Replace the polygon driver.
Remove the polygon driver with the shield case attached.
(7) Disconnect the connector (CN1) on the LDD08A board.
(8) Remove the top cover from the scanning optics unit.
(9) Remove the polygon cover.
(10) Remove the polygonal mirror assembly.
(9)
Polygon cover
(10)
Polygonal mirror assembly
(Pay attention to its
mounting direction.)
Polygon mirror face
009-051-04
009-051-06
04.20.2001 FM3006
05.31.99 FM2460
3
4
5
6
Screw x4
Screw x4
(7) LDD08A CN1
CR-IR341 Service Manual
7
8
9
10
(8)
Top cover
Screw x7
Polygon mounting direction
Polygon
Filter
(4)
Clamp
BR M4x8
MC - 154
Removal"
Cable
The narrower edge
F
R
width should face
O
this direction.
mirror
N
face
T
(5)
Tie
Filter
(6)
Polygon driver
Screw x2
MC - 154
FR1H1170.EPS