C
11
12
ONTENTS
To turn HV ON
When the HV switch is in the OFF position, menu item 9 is displayed as "9.HV ON".
0.QUIT
1.ERROR LOG UTILITY
2.CONFIGURATION SETTING
3.TEST MODE
4.ELECTRICAL UTILITY
5.SCANNER UTILITY
6.MECHANICAL UTILITY
7.FILE UTILITY
8.BACKUP MEMORY
9.HV ON
10.IP UTILITY
>
(1) Select "9. HV ON".
When the high-voltage switch is set to ON, the following message appears.
TURNING ON THE HV IS IN PROGRESS.
RESULT - OK
Meaning: The HV is now being turned ON.
If any error occurs, either of the following messages is displayed.
HV ERROR
ANALOG POWER SUPPLY ERROR
HV OFF
(2) Select "0. QUIT".
009-051-05
009-051-05
11.20.99 FM2638 (1)
11.20.99 FM2638
13
14
15
16
CR-IR341 Service Manual
17
Meaning: High-voltage power supply failure
Meaning: Analog power supply failure
Meaning: HV switch OFF
MU - 129
MU - 129