FujiFilm CR-IR341 Service Manual page 672

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
(8) Attach the two take-out screws to the light-collecting mirror.
They facilitate grasping the light-collecting mirror to prevent it from dropping into a gap.
(9) Remove the two screws that retain the light-collecting mirror in place.
(10) Remove the light-collecting mirror.
Grasp the take-out screws and remove the light-collecting mirror while using care to
avoid contact with surrounding objects.
Reflection
face
009-051-04
05.31.99 FM2460
3
4
5
6
(9), (10)
Light-collecting mirror
BR M4x8, x2
CR-IR341 Service Manual
7
8
9
10
(8)
Take-out screw
M3
MC - 181
FR1H1541.EPS
MC - 181