FujiFilm CR-IR341 Service Manual page 777

Fuji computed radiography
Table of Contents

Advertisement

C
11
12
ONTENTS
16.10 CPU90F Board
CAUTION
When servicing any printed circuit board, be sure to wear an anti-static wristband to ensure
proper grounding. If not, static electricity built on the human body may cause damage to electric
parts mounted on the board.
NOTE
With board version N or later, the software version of the machine should be updated to A09 or
later. Earlier software versions do not support a newer LSI (discontinuation supported) on the
daughter board (ETH90F board) of the CPU90F board.
Setting CPU90F Board
Make sure that the CPU90F board is set as follows.
CNSIO: No short pin (open)
T1, T2: Nos. 2 and 3 are shorted.
DIS1:
DIS2:
SW1:
CNSI0
009-051-06
009-051-04
04.20.2001 FM3006
05.31.99 FM2460
13
14
15
16
Nos. 1, 3, and 4 are OFF; only No. 2 is ON.
All OFF
"6"
No short pin
(OPEN)
CN2
ETH90F board
T4
NFB2
DIS1
OFF position
4
3
2
1
ON position
CR-IR341 Service Manual
DIS2
OFF position
4
3
2
1
ON position
MC - 282
T1,2
Short pin
3
2
1
CN1
NFB1
SW1
Set to "6"
FR1B1136.EPS