10.11 Checking and Replacing SUS Belt/Rubber Belt/Tensioner
CAUTIONS
• Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
• Wear gloves during checks and replacement, because the SUS belt may be soiled or dam-
aged.
Checks
(1) Remove the following covers.
Lower front cover, lower right-hand side cover
"3. Removing and Reinstalling Covers"
(2) Remove the transparent cover located on the right-hand side.
Using a blower, remove dust deposited on the transparent cover.
(3) While rotating the drive shaft, check the SUS belt, rubber belt, pulley, and tensioner for
any soil or flaw.
• If any soil is found, clean it off.
SUS belt:
Rubber belt: Clean any soil off with a moistened cloth.
• If any flaw is found, replace the SUS belt or rubber belt.
Subscanning
unit
(4) Reinstall the right-hand side trsnsparent cover.
009-051-04
05.31.99 FM2460
3
4
5
6
Clean any soil off with an ethanol-moistened cloth.
Subscanning unit
CR-IR341 Service Manual
7
8
9
10
(2)
Tensioner
SUS belt
Drive
shaft
MC - 192
Rubber belt
Pulley
Right-hand side transparent cover
BR M4x8, x2
FR1H1189.EPS
MC - 192