FujiFilm CR-IR341 Service Manual page 689

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
Replacement of flywheel
(1) Pull out the scanner unit.
"7. Pulling Out and Pushing in Scanner Unit"
(2) Remove the light-collecting guide assembly.
"10.2 Servicing Light-Collecting Guide"
(3) Remove the after-reading conveyor.
"11.1 Removing and Reinstalling After-Reading Conveyor"
(4) Remove the transparent cover located on the left-hand side.
(5) Remove the spring.
(6) Remove the Kapton
(7) Insert an Allen wrench or so forth into the hole of the shaft located inside the side plate
to secure it in place, and remove the nut that retains the flywheel.
(8) Replace the flywheel.
(5)
Spring
(9) Install a new flywheel by reversing the removal steps.
NOTE
After installing the flywheel, rotate it by hand to ensure that the Kapton
come off the pulley.
(10) Power ON the machine.
(11) Check MZ1.
Make sure that the motor rotates normally.
"10.8 Checking and Replacing Subscanning Motor (MZ1)"
(12) Check image performance.
"3.2 Checking Image" in Preventive Maintenance Volume
009-051-04
05.31.99 FM2460
3
4
5
6
®
belt.
(6)
Kapton
(4)
Left-hand side
transparent cover
BR M4x8, x2
CR-IR341 Service Manual
7
8
9
10
Side plate
Hole
Shaft
Allen wrench
(7)
(8)
Nut
Flywheel
¤
belt
FR1H1817.EPS
MC - 198
®
belt does not
MC - 198