FujiFilm CR-IR341 Service Manual page 557

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
5.4
Checking and Adjusting Inch/Metric Sensor (SB4)
Checks
If anything abnormal is found during checks, replace SB4.
"2.2 Checking Various Sensors"
Replacement
(1) Remove the bracket that retains SB4 in place.
NOTE
Remove the bracket before disconnecting the connector (SB4). If you attempt to
disconnect the connector with the bracket fixed, moment is applied to the connector,
thus damaging the connector.
(2) Disconnect the connector (SB4).
(3) Remove the bracket from SB4.
(2)
Connector
Pan-head M3x8, x2
(4) For reinstallation, reverse the removal steps.
NOTES
• When mounting connector SB4 on the bracket, ensure that the SB4 lower surface is
flush with the bracket cut surface.
Flush
• When mounting the bracket in the up-down IP removal unit, screw it down while
pressing it downward.
(5) Check SB4.
" Checks"
009-051-04
009-051-04
05.31.99 FM2460 (1)
05.31.99 FM2460
3
4
5
6
(1)
(3)
Bracket
SB4
Sems M3x6, x2
Bracket
SB4
CR-IR341 Service Manual
7
8
9
10
FR1H4015.EPS
MC - 66
FR1H1078.EPS
MC - 66