5.7
Replacing IP Leak Valve (SVB1)
(1) Rotate the timing belt until the IP removal unit is moved up (away about 250 mm from
the tray of the cassette set unit) to facilitate SVB1 removal.
(2) Disconnect the connector (CNSVB1).
(3) Cut the cable tie that retains the hose in place.
(4) Pull out the hose from SVB1.
(5) Remove SVB1.
Remove the two SVB1 retaining screws behind the machine, and take out SVB1.
(6) For reinstallation, reverse the removal steps.
009-051-04
05.31.99 FM2460
3
4
5
6
IP removal unit
(1)
Timing belt
Cut cable tie, x2
CNSVB1
CR-IR341 Service Manual
7
8
9
10
Up-down IP removal unit
(3)
(2)
(4)
Hose
(5)
SVB1
BR M4x8, x2
MC - 70
FR1H1007.EPS
MC - 70