FujiFilm CR-IR341 Service Manual page 822

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
Item
IDT
Item
number
type
41
EM1
41
EM2
41
EM3
41
EM4
41
EM5
U_LP
42
43
ID_AP
44
ID_DST
45
UID_STI
DEF_
46
DENSITY
47
JSTFY
009-051-06
009-051-05
04.20.2001 FM3006
11.20.99 FM2638
3
4
5
6
Initial value
Setup description
Menu code for executing backup
Space
when EBK = 1
Menu code for executing backup
Space
when EBK = 1
Menu code for executing backup
Space
when EBK = 1
Menu code for executing backup
Space
when EBK = 1
Menu code for executing backup
Space
when EBK = 1
OEM printer setting
1
Image reversal mode (orientation)
0
setting range
0
Distribution code setting
Setting of study instance UID type
Setting used to issue study
information for QA-DICOM output.
3
Used to customize, on a hospital-
by-hospital basis, how to bind
images per study on the QA side
Default setting of image
processing density
Used to designate the machine s
default processing density.
This setting is effective only when
a setting other than 0 or 1 is
registered from IDT3 or the like
(in cases where IDT3, which
0
cannot designate the processing
density, is connected via IDT4,
or where the processing density
is not designated due to a software
bug of IDT4).
Normally, the value that is set on
the IDT side or by the internal
menu of the CSL type is effective,
so its initial value need not be
changed.
00010000
Definition of right or left justification
00001111
for long format
CR-IR341 Service Manual
7
8
9
10
Four hexadecimal digits are used
for menu code designation.
Four hexadecimal digits are used
for menu code designation.
Four hexadecimal digits are used
for menu code designation.
Four hexadecimal digits are used
for menu code designation.
Four hexadecimal digits are used
for menu code designation.
0: OFF
1: ON
0
:
L-R reversal
1
:
T-B reversal
2
L-R,T-B reversal
:
3
:
L-R reversal/90-deg.Rotation
4
:
L-R,T-B reversal/90-deg.Rotation
0
DISABLE
:
1
:
ENABLE
0: <org root>.2.0.<Private Image UID'>
(IDT Type)
0: <org root>.2.1.<MAC Address>.
<Time Stamp>.<Unique ID>
(CSL Type)
1: <org root>.2.1.<Accesion Number'>
2: <org root>.2.2
.<Study Date + Modality(CR)>
.<Accesion Number'>
3: <org root>.2.3
.<Study Date + Modality(CR)>
.<Patient ID'>
4: <org root>.2.4
.<Study Date> .<Patient ID'2>
.<Requesting Service'>
0
Standard Density
:
1
High Quality Density
:
<Variable part>
0: Left justified; 1: Right justified
<Corresponding to each digit
of the variable part>
a: Patient ID
b: Patient name
c: Exposure menu name
d: Examination number
e: Department name (alphanumeric)
f-l: Reserved (default: left justified)
m-p: Reserved (default: right justified)
MU - 39
Parameters
Remarks
LENGTH
LENGTH
LENGTH
LENGTH
LENGTH
No change
required.
<org root>=
1.2.392
.200036.9125
Common to
DMS output,
CRT display,
and FILM
output.
MU - 39
=4
=4
=4
=4
=4
TR1B1009.EPS

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents