13E1 Feed IP Suction Error
Cause
To convey the IP in the cassette, PB1 is turned ON to suck the IP with the suction
cups, but SB5 does not turn ON.
Supplementary explanation
To suck the IP in the cassette, the suction cups are moved from the home position
toward the cassette by the driving of MB1.
To suck the IP with the suction cups, PB1 is turned ON.
When the suction of the IP by the suction cups is detected, SB5 is turned ON.
This error occurs when SB5 does not turn ON during the above operation.
Sensor related
Error-related I/O locations
Error-related system block diagram
AC IN
DC OUT2
JPS-1
DC OUT3
009-051-04
05.31.99 FM2460
3
4
5
6
ON: Close, OFF: Open
PB1
IP suction
SNS08A
CN
CN13
CN4
DRV03D
CN3
CN3
CNB4
DRV08A
CN1
CR-IR341 Service Manual
SB5
Suction sensor
FR1H0125.EPS
CNB1
CNB2
CNPB1
PB1
FR1H0126.EPS
MT - 156