Replacement
(1) Remove the mounting bracket together with SA3, 7, 11, 15.
(2) Disconnect the connectors (SA3, 7, 11, 15).
(3) Remove SA3, 7, 11, 15 from the mouting bracket.
The sensor is secured to the bracket by four latches. To remove the sensor, pinch the
latches by fingers.
SA3
SA7
SA11
SA15
(4) Install a new sensor.
"3.5 Sensors Not Requiring Adjustment of Its Mounting Position" in Machine
Description
(5) Connect the connectors (SA3, 7, 11, 15).
(6) Attach the mounting bracket.
Screw down the mounting bracket while pushing it upward.
(7) Check the sensor replaced.
" Checks"
009-051-04
05.31.99 FM2460
3
4
5
6
FR1H1056.EPS
CR-IR341 Service Manual
7
8
9
10
MC - 39
MC - 39