FujiFilm CR-IR341 Service Manual page 712

Fuji computed radiography
Table of Contents

Advertisement

C
11
12
ONTENTS
Reinstallation
(1) Install the erasure conveyor along the shock absorber-attached cover into the ma-
chine.
Align the erasure conveyor against the positioning bracket located on the left side of
the machine (into the machine as seen from the direction of insertion), and secure it in
place.
(2) Attach the timing belt onto the timing belt wheel.
(3) Put the tensioner back where it was.
Loosen the single screw that retains the tensioner. Flick the timing belt several times to
ensure that it snugly fits onto the belt wheel, and then tighten the screw on the
tensioner in place where appropriate tension is attained.
Positioning bracket
Shock absorber-
attached cover
(4) Connect the connectors (CNF1, 2, CNSOLF1 to the erasure conveyor; CNC1 to the
before-side-positioning conveyor).
CNF1
009-051-04
05.31.99 FM2460
13
14
15
16
CNSOLF1
Erasure conveyor
CR-IR341 Service Manual
Erasure conveyor
(2)
Timing belt
Before-side-positioning
Erasure conveyor
conveyor
(1)
(1)
Erasure conveyor
Spacer x2
BR M4x8, x2
CNF2
Before-side-positioning
conveyor
MC - 221
BR M4x8
(3)
Tensioner
CNC1
FR1H1704.EPS
CNC1
FR1H1518.EPS
MC - 221