(16) Make sure that there is a gap between the upper rubber roller and lower rubber roller.
If there is no gap, make adjustment as appropriate by rotating the adjustment screw.
10.15 Replacing Guide Plate of Subscanning Unit
(1) Remove the rubber roller located above the subscanning unit.
"10.13 Replacing Rubber Roller Located Above Subscanning Unit"
(2) Remove the light-collecting mirror.
"10.4 Servicing Light-Collecting Mirror"
(3) Remove the right-hand side transparent cover.
(4) Remove the two screws that retain the grip arm located on the right-hand side of the
machine.
Grip arm
Subscanning unit
009-051-06
009-051-04
04.20.2001 FM3006
05.31.99 FM2460
3
4
5
6
Grip arm
Adjustment screw
Gap (reference value): 0<A–F≤0.3 (mm)
Subscanning unit
(3)
Right-hand side transparent cover
BR M4x8, x2
CR-IR341 Service Manual
7
8
9
10
Adjustment screw
A
B
Lower rubber roller
(4)
Grip arm retaining screw
BR M3x6, x2
MC - 202
D
Upper rubber roller
E
C
Subscanning unit
MC - 202
F
FR1H6521.EPS
FR1H1529.EPS