ONTENTS
No. 203
Improper image (3)
• Although IP conveyance is normal and the film
characters are outputted normally, the image and
border are not generated.
Improper image (4)
No. 204
• A full image in each size of 14" x 17" (35cm x
43cm), 14" x 14" (35cm x 35cm), and 10" x 12"
(18cm x 24cm), as well as a two-in-one image of
8" x 10" (18cm x 24cm) size, is disturbed.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
<Probable Cause>
• The IMG08A board is faulty.
<Check and Reference>
• Reset the machine, or power OFF the machine
and then back ON again, to perform self-
diagnostics during machine initialization. If the
results of the self-diagnostics indicate an error,
replace the relevant board.
<Probable Cause>
• The contents of the memory containing
enlargement/reduction ratios in the IMG08A
board is corrupted.
<Check and Reference>
• Reset the machine, or power OFF the machine
and then back ON again, to perform self-
diagnostics during machine initialization. If the
results of the self-diagnostics indicate an error,
replace the relevant board.