Preliminary steps:
1. Power OFF the machine and remove the covers.
2. Turn OFF the high-voltage switch (S1) of the SCN08A board.
3. Check for any foreign matter (lead character marking chips, etc.) inside the machine, and if any, remove it all.
Power OFF
Is the IP normal?
Make sure that it is not curled
or bent.
Make sure that the barcode
display portion is free of soil.
Y
Power OFF
Is the BCR reading face normal?
Make sure that the BCR reading
face is free of soil.
Y
Power ON
Is the BCR reading face is lit?
Make sure that the BCR reading
face is lit in red.
Y
Restart the machine
and check its status.
009-051-04
05.31.99 FM2460
3
4
5
6
N
Replace the IP.
N
Clean the BCR reading face.
Power OFF
N
Are the connector
connections from BCR to
SNS08A board normal?
Replace the following
parts in order named.
BCR
SNS08A board
CR-IR341 Service Manual
N
Check the error-related
system block diagram and
connect the connectors
Y
as appropriate.
MT - 143
FR1H0115.EPS