13.5 Replacing the Monitor
WARNINGS
• When replacing the monitor, be sure to turn OFF the circuit breaker. If you replace the monitor
with the circuit breaker ON, you could receive a shock.
• Since the monitor contains high-voltage generating parts (see the figure below), never touch
them. If any foreign matter enters the monitor, be sure to remove it. If it is allowed to stay, a
risk of fire or electric shock will be caused.
: Never touch the shaded portions.
Remove the upper left-hand side cover, upper right-hand side cover, and front cover.
(1) Remove the power switch cover.
Even when the power switch cover is removed, the connectors are still connected.
(2) Disconnect connectors LED08A CN1 and CN2.
(3) Open the bracket after removing two screws.
009-051-04
05.31.99 FM2460
13
14
15
16
Board (700V max.)
Board
Deflection yoke coil
(250V max.)
(2)
LED08A CN1, 2
(1)
Power switch cover
BR M4x8, x2
CR-IR341 Service Manual
Anode cap
Flyback transformer
Board (700V max.)
Anode lead
(approx. 12kV)
FR1H4022.EPS
(3)
Bracket
BR M4x8, x2
FR1H4020.EPS