Accelerated phosphorescence light emitted from a latent image due to laser beam irradiated
onto the latent image on the IP surface is collected by the light-collecting mirror, and the
excited light (660 nm laser light) is cut off by the filter mounted on the photomultiplier
section, and the resulting light is guided to the photomultiplier (PMT).
The dust-shield guide prevents dust deposits onto an area between the light-collecting
guide/mirror section and scanning optics unit.
Light-collecting
Accelerated Phosphorescence Light
009-051-04
05.31.99 FM2460
3
4
5
6
Laser light
guide
IP surface
CR-IR341 Service Manual
7
8
9
10
Light-collecting mirror
Accelerated phosphorescence light
MD - 55