No. 304
Improper format (1)
• The right or left edge of the image is missing.
Improper format (2)
No. 305
• The bottom or top edge of the image is missing.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
<Probable Causes>
• The position of the scanning optics unit is
improper (in the main-scan direction).
• The side-positioning operation is improper.
• The irradiation field is improper.
• The parameter for adjusting the main-scan format
is improper.
<Check and Reference>
<Probable Causes>
• The IP feed by FFM (MZ1) is improper when the
IP is fed into the subscanning unit.
• The IP leading-edge sensor (SZ1) is faulty.
• The rpm of the FFM (MZ1) is improper.
• The irradiation field is improper.
<Check and Reference>
MT - 190