[6] Set Sensitivity Correction
NOTE
When set sensitivity correction is performed such that sensitivities of several readers are
adjusted to a consistent level, the same IP should be used. It should also be verified in advance
that there is no variation in dosage when several exposures are performed with the same
output. If the dosage varies from one exposure to another, set sensitivity correction cannot be
accomplished.
After set sensitivity correction, perform the steps described in "[5] Checks on Correction
Results" again.
(1) Uniformly expose a 14" x 14" (35cm x 35cm) or 14" x 17" (35cm x 43cm) IP at 0.5 to
9.99 mR.
(2) Perform recording under SENSITIVITY of the IDT's Test Menu, and generate image
output.
NOTE
Image output should be generated 10 minutes after the uniform exposure of the IP.
(3) After checking the film output, take out the IP used for shading correction.
(4) Select "5. SCANNER UTILITY", "7. SHADING/SENSITIVITY" and "4. SENSITIVITY
DATA" in sequence.
(5) Enter the current value and a desired correction value in sequence.
The correction value is set.
(6) Generate film output and confirm the correction result.
(7) Select "5. SCANNER UTILITY", "8. DATA MANAGEMENT" and "2. SAVE SENSITIVITY
DATA" in sequence.
The sensitivity data after correction is saved to the hard disk.
(8) Exit M-Utility.
009-051-05
11.20.99 FM2638
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
MC - 142
MC - 142