FujiFilm CR-IR341 Service Manual page 553

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
5.2
Checking and Replacing Cassette Inlet IP Sensor (SB2)
Checks
CAUTION
Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
(1) Power ON the machine.
The initialization starts, and after an elapse of about 30 to 60 seconds, "the soft-
ware version number (Axx)" appear on the operation panel.
(2) Immediately after the software version number (Axx) appears, sequentially touch the
operation panel upper left corner and upper right corner.
M-Utility is activated.
When the screen reads STEP95, the operation panel is inoperative and does not
respond at all even if you touch it. In such an instance, perform a reset and wait until
the software version number (Axx) appears.
(3) Select "5. MECHANICAL UTILITY", "1. INITIALIZE" and "4. SENSOR" in sequence.
(4) Select "2. MONITOR".
A list of sensor Nos. is displayed.
(5) Type in the number of the cassette inlet IP sensor (SB2).
The machine is now in the sensor monitoring enabled state.
(6) Using a slip of paper or the like, block the sensor light path.
A message indicating that the sensor is CLOSE is displayed on the operation panel.
(7) Remove the slip of paper or the like that has blocked the sensor light path.
A message indicating that the sensor is OPEN is displayed on the operation panel.
(8) Repeat steps (6) and (7) several times.
If anything abnormal is found, such as no message displayed, then replace the sensor.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
IP removal unit
MC - 62
(6), (7)
SB2
Slip
of paper
FR1H1293.EPS
MC - 62