FujiFilm CR-IR341 Service Manual page 212

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
(7) Detailed Sequence of Driving Shaft Self-Diagnostics
Fig. 5.9.7 Driving Shaft Self-Diagnostics
Preprocessing: Fig. 5.9.1
START
Speed data setting
Driven shaft grip release driving
DRIVE (MZ2/ST speed ON)
TZ25 STRAT
Timer START
Is SZ2 opened within TZ25?
N
Y
Driven-side grip release stop
DRIVE (MZ2/OFF) (hard)
Timer STOP
Timer>TZ27 or Timer< TZ26
N
Y
Rotation-speed correction
Sequence flow chart
See 5.9 (8).
Speed data update
DELAY TZ51
Speed data setting
END
Table 5.9.7 Operation Parameters
Symbol
Driven shaft grip timeout (self-diagnostics)
TZ25
Lower limit for driven shaft grip spec (self-diagnostics)
TZ26
Upper limit for driven shaft grip spec (self-diagnostics)
TZ27
Wait for initialization grip operation
TZ51
009-051-04
05.31.99 FM2460
3
4
5
6
The previous latest data
(corrected) in COMS is used.
03CA
Description
TR1H0086.EPS
CR-IR341 Service Manual
7
8
9
10
Driven shaft grip release stop
DRIVE (MZ2/OFF) (soft)
Level 0 error
Driving shaft grip release
self-diagnostics error
FR1H0436.EPS
Value
5 sec
2.2 sec
2.4 sec
0.5 sec
MD - 149
MD - 149

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents