Replacement
(1) Remove the side-positioning conveyor.
"8.1 Removing and Reinstalling Side-Positioning Conveyor"
(2) Remove the E-ring.
(3) Take out the triangular-shaped cam from the shaft.
(4) Remove the wire and shaft from the triangular-shaped cam.
(4)
Nut
(5) Install the wire and shaft removed at step (4) to a new triangular-shaped cam.
NOTE
When installing the triangular-shaped cam to the shaft, use care not to confuse the
orientation of the triangular-shaped cam.
(6) Reinstall the triangular-shaped cam to the shaft.
(7) Reinstall the E-ring.
(8) Reinstall the side-positioning conveyor.
"8.1 Removing and Reinstalling Side-Positioning Conveyor"
(9) Check the grip mechanism of the side-positioning conveyor.
" Checks"
009-051-04
05.31.99 FM2460
3
4
5
6
(3)
Shaft
Triangular-shaped cam
(4)
Shaft
(4)
Wire
FRONT
CR-IR341 Service Manual
7
8
9
10
(2)
E-ring
Truss M3x6
FR1H1784.EPS
FRONT
MC - 123
FR1H1785.EPS
MC - 123