FujiFilm CR-IR341 Service Manual page 545

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
(5) Loosen the single screw that retains the 24x30 mammo rearward toppling prevention
bracket.
(6) Make adjustment as appropriate so that the descent of the trailing edge of the movable
roll is 1 mm or less.
By sliding the rearward toppling prevention bracket in the direction of arrow as shown
below (toward the shutter assembly), the descent of the trailing edge of the movable
roll is decreased.
Top view
(7)
BR M4x8
Side view
(7) Reinstall the shutter assembly.
"4.3 Removing and Reinstalling Shutter Assembly"
(8) Reinstall the cassette set unit.
"4.2 Removing the Reinstalling Cassette Set Unit"
(9) Push in the tray of the cassette set unit.
"4.1 Pulling Out and Pushing In Tray of Cassette Set Unit"
009-051-04
05.31.99 FM2460
3
4
5
6
Trailing edge of movable roll
24x30 mammo movable roll
(8)
24x30 mammo rearward toppling
prevention bracket
The descent is decreased.
Slightly push the trailing edge of the movable roll.
Movable roll
CR-IR341 Service Manual
7
8
9
10
FR1H1765.EPS
Movable roll
Ensure that it is 1 mm or less.
Movable roll support guide
MC - 54
FR1H1766.EPS