10.7 Checking and Replacing Driven-Side Grip Release HP
Sensor (SZ3)
Checks
If anything abnormal is found during checks, replace SZ3.
"2.2 Checking Various Sensors"
NOTE
When monitoring SZ3, insert an Allen wrench into the cam of MZ3 and rotate it clockwise to
check the OPEN/CLOSE status of SZ3.
Replacement
(1) Pull out the scanner unit.
"7. Pulling Out and Pushing In Scanner Unit"
(2) Remove the light-collecting guide assembly.
"10.2 Servicing Light-Collecting Guide"
(3) Remove the side-positioning conveyor.
"8.1 Removing and Reinstalling Side-Positioning Conveyor"
(4) Disconnect the connector (SZ3).
(5) Remove the SZ3 bracket.
(6) Take out SZ3 from the bracket.
Subscanning unit
(7) Reverse the removal steps to reinstall a new SZ3.
(8) Check SZ3.
" Checks"
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
Left side of
subscanning unit
Rear side plate
(5)
Bracket
Sems M3x6
MC - 187
(4)
(6)
Connector
SZ3
FR1H1288.EPS
MC - 187