Reinstallation
For reinstallation, reverse the removal steps.
CAUTIONS
• Before inserting the cassette set unit into its position, place the IP removal arm in the home
position. Also, adjust the IP removal height for the home position or the first to fourth shelf
position. If an attempt is made to insert the cassette set unit with the up-down IP removal
assembly positioned elsewhere, the machine may be damaged.
• When placing the cassette set unit in its position, insert it along the positioning brackets
because the sensor may be damaged.
• When installing the cassette set unit, properly position it by pressing it against the positioning
brackets. Screw down the cassette set unit while pushing its upper end toward the up-down
IP removal unit.
Cassette set unit
BR M4x8, x6
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
Top view
Push top stay
of up-down IP
removal unit
FR1H1422.EPS
MC - 29
Positioning bracket
Cassette set unit
Top stay of
up-down IP
removal unit
Cassette set unit
MC - 29
F
R
O
N
T
FR1H1051.EPS