FujiFilm CR-IR341 Service Manual page 917

Fuji computed radiography
Table of Contents

Advertisement

C
11
12
ONTENTS
13. 8. BACKUP MEMORY: Backing Up Memory
NOTE
Be sure to reset the machine after clearing the backup memory. Unless the machine is reset,
the backup memory is not cleared.
It is used when the backup memory is cleared.
(1) Select "1. CLEAR".
The following message appears.
ARE YOU SURE TO CLEAR ALL THE BACKUP MEMORIES?
1.YES 2.NO (DEFAULT=2) :
Meaning: Are you sure to initialize the backup memory?
(2) Select "1".
The backup memory is initialized, with the following message displayed.
BACKUP MEMORIES ARE CLEARED.
Meaning: The backup memory is initialized.
(3) Repeat "0. ENT".
M-Utility is exited.
(4) Press the RESET switch of the machine.
The following files are cleared.
• IMG
Set processing information, information for image number generation, HV ON/OFF
information
• IPH
IP position information, subscanning grip correction information
• PNL
Audible click tone enabled/disabled, parallax correction data
• DST
Output LP information (NET/LOCAL)
• JNL
Error log information
• LIF
Format frequency information
• MFC
Setup options information
• CSL (for CSL specification only)
Multiframe-related information, set processing-related information, menu selection
information