[3] Input of Read Width Adjustment Value
(1) Activate M-Utility.
" Activating and Exiting M-Utility"
(2) Select "5. SCANNER UTILITY", "6. FORMAT" and "2. FREQ ADJUST" in sequence.
(3) Enter the read width adjustment value and press the [ENT] key.
The read width is adjusted.
Type in the value indicated on the label located on the side of a new scanning optics
unit.
Scanning optics unit
(4) Select "5. SCANNER UTILITY", "8. DATA MANAGEMENT" and "3. SAVE FORMAT
DATA" in sequence.
The read width adjustment value so adjusted is saved in the hard disk.
[4] Input of Read Start Position Adjustment Value
(1) Uniformly expose a 14" x 14" (35cm x 35cm) or 14" x 17" (35cm x 43cm) IP at about
10 mR.
Note that figures in parentheses are applicable when the machine is set in metric unit
of measure.
(2) Perform recording under SENSITIVITY of the IDT's Test Menu, and generate image
output.
(3) Measure the white blank portion of the output film, in steps of 2 mm, from the outer-
most corner of the image frame.
For measurement, use the steel rule or graduated loupe.
White blank
009-051-05
11.20.99 FM2638
3
4
5
6
FR1H1101.EPS
2mm or less
Image frame
FR1H1102.EPS