1
2
C
ONTENTS
13AE No MPM Code Available
Cause
An attempt is made to perform image processing with the information of the patient
exposed, but no relevant image processing conditions of the machine are found.
Supplementary explanation
The exposure information of the patient (exposure anatomical region, exposure
menu, modality) is inputted to the ID terminal via an ID card.
During IP reading conveyance, the barcode of the IP is read by the BCR of the side-
positioning conveyor to search for image processing corresponding to the exposure
information of the patient.
This error occurs when image processing corresponding to the exposure information
of the patient is not found.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
MT - 152
MT - 152