(7) Type in the number of the sensor to be checked (SA2, 6, 10, 14).
The machine is now in the sensor monitoring enabled state.
(8) Loosen the screw on the mounting bracket of the sensor to be checked (SA2, 6, 10,
14).
(9) Secure the manual release lever of the cassette set unit in place with a cable tie, so
that the hold pin is withdrawn.
(10) While pushing down the latch by hand, set a scale in such a position that the distance
from the end face of the plate to the tip of the scale is 1.5 mm.
(11) With the tip of the scale in contact with the actuator of the sensor (SA2, 6, 10, 14) to
be checked, slowly move the mounting bracket bit by bit toward the scale and tighten
the screw in place where the sensor changes status from OPEN to CLOSE.
(12) Slide the scale until the distance between the end of the plate and the tip of the rule is
1.0 mm, and make sure that the sensor (SA2, 6, 10, 14) transitions to OPEN.
(10)
(13) Repeat steps (6) through (12) to adjust the sensor mounting position.
(14) Cut the cable tie that was attached at step (9).
(15) Select "0. QUIT".
The screen switches back to "5. MECHANICAL UTILITY" menu.
(16) Select "0. QUIT".
The screen returns to the main menu.
(17) Power OFF the machine.
(18) Check the sensor adjusted.
" Checks"
009-051-06
009-051-04
04.20.2001 FM3006
05.31.99 FM2460
3
4
5
6
(10)
Latch
Scale
CR-IR341 Service Manual
7
8
9
10
Plate
SA2
(10)
Scale
SA6
1.5mm : Close
SA10
1mm
: Open
SA14
MC - 37
Actuator
SA2, 6, 10, 14
(8), (11)
Mounting bracket
BR M4x8
End of calipers
Plate end face
FR1H1412.EPS
MC - 37