FujiFilm CR-IR341 Service Manual page 649

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
(17) Touch the U-Utility button.
The screen switches to the U-Utility mode.
(18) Touch the upper left and upper right corners of the operation panel in sequence.
M-Utility is activated.
(19) Select "5. SCANNER UTILITY", "3. LASER", and "2. ON" in sequence.
(20) Select "5. SCANNER UTILITY", "2. POLYGON", and "2. ON" in sequence.
(21) Check if D14 of the SCN08A board is lit (blinking quickly).
• If it is lit, set "LASER" and "POLYGON" to OFF, and restart the machine.
If the machine starts up normally, the SYN08A board is normal.
If the machine does not start up normally, replace the SYN08A board.
• If it is not lit or is blinking (blinking but not quickly), replace the SYN08A board.
009-051-04
05.31.99 FM2460
3
4
5
6
S1
D14
SCN08A board
FR101105.EPS
CR-IR341 Service Manual
7
8
9
10
MC - 158
MC - 158