11.3 Replacing Grip Release Motor (ME1)
The grip release motor (ME1) must be replaced with the lower front cover and lower front
inner cover removed. It is not necessary to pull out the scanner unit.
(1) Remove the cable clamp.
(2) Disconnect the connector (CNME1).
(3) Remove ME1.
Remove the two screws that retain the front side of the ME1. Also remove the single
screw that retains the stay on the inner side of the ME1 by accessing it from the side
plate.
(4) Remove the stay and timing belt wheel from ME1.
After-reading
(3)
BR M4x8
(5) Reverse the removal steps to reinstall a new ME1.
009-051-04
05.31.99 FM2460
13
14
15
16
conveyor
(4)
Stay
Pan M4x10
(4)
Timing belt wheel
(3)
Pan M4x10 x2
CR-IR341 Service Manual
(2)
CNME1
ME1
(3)
ME1
FR1H1128.EPS