FujiFilm CR-IR341 Service Manual page 140

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
(4) Sequence of Side-Positioning
The IP conveyed to the side-positioning stopper is side-positioned.
The operation of the side-positioning stopper is performed during the sequence of side-
positioning conveyance described in the preceding subsection 2.6 (3).
Right-hand side
plate of side-
positioning
conveyor
Side guide
(out)
Side-Positioning Mechanism and Driving Section
Latch
assembly
FRONT
Belt
MD1
Side-positioning mechanism
Table 2.6.4a Operation Parameters
Symbol
PD21
Conveyance grip release
PD24
Conveyance grip (slow)
TD11
Wait for settlement of IP that is not side-positioned
TD12
Wait for side-positioning return
TD13
Wait for re-side-positioning
TD14
Wait for settlement of IP that has been side-positioned
Table 2.6.4b Operation Parameters
Symbol
PD12
Side-positioning (stopper
PD13
Side-positioning (side-positioning 1
PD14
Side-positioning (back
PD15
Side-positioning (side-positioning 2
009-051-04
05.31.99 FM2460
3
4
5
6
Fig. 2.6.4 Side-Positioning
Preprocessing: Fig. 2.6.1, Fig. 2.6.2, Fig. 3.2.1, and Fig. 5.6.1
The IP conveyed
from the before-
side-positioning
conveyor to the
side-positioning
FR1H0338.EPS
conveyor is
side-positioned.
Sensor arm
for SD1
SD1
FR1H0339.EPS
Description
Description
side-positioning 1)
back)
side-positioning 2)
HP)
CR-IR341 Service Manual
7
8
9
10
START
Side-positioning conveyance grip release
DRIVE (MD2/PD21)
DELAY TD11
Side-positioning
(stopper
side-positioning 1)
DRIVE (MD1/PD12)
DELAY TD12
Side-positioning
(side-positioning 1
DRIVE (MD1/PD13)
DELAY TD13
Side-positioning
(back
side-positioning 2)
DRIVE (MD1/PD14)
DELAY TD14
Side-positioning conveyance grip (slow)
DRIVE (MD2/PD24)
Side-positioning (side-positioning 2
DRIVE (MD1/PD15)
Is SD1 closed?
N
Y
END
Value
256 p
144 p
0.2 sec
0.1 sec
0.1 sec
0.2 sec
TR1H0027.EPS
14x17 I
14x14 I
10x12 I
8x10 I
18x24 ST
98 p
98 p
778 p
778 p
872 p
68 p
68 p
68 p
68 p
-68 p
-68 p
-68 p
-68 p
-68 p
162 p
162 p
842 p
842 p
936 p
The side-positioning
operation is performed
twice in order to increase
positional accuracy.
The second side-positioning
back)
movement is so small that it is
difficult to visually check that
movement.
HP)
Level 2 error
23BC
Side-positioning operation
error
Side-positioning
HP positioning
Fig.2.6.5
OK
24x30 ST
18x24 HR
24x30 HR
872 p
840 p
440 p
68 p
68 p
34 p
34 p
-68 p
-34 p
-34 p
936 p
904 p
504 p
TR1H0028.EPS
MD - 77
FR1H0337.EPS
MD - 77