FujiFilm CR-IR341 Service Manual page 447

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
3.
Troubleshooting Procedures Based on
Abnormal Images
For image troubleshooting, it is necessary to identify a trouble spot in the machine, judging
from the characteristics of the image. In an image artifact or abnormality occurs, perform
troubleshooting in accordance with the basic analytical flow described below.
Note, however, that troubleshooting procedures for the image output system and X-ray
exposure system are omitted in this chapter.
Failure in
scanner unit
Troubleshoot
the scanner unit.
"3.1 Image
Troubleshooting
for Scanner Unit"
3.1
Troubleshooting for Scanner-Related Errors
Procedures
(1) Check the error log for the time when the abnormal image occurred.
View the error log to check for any error whose occurrence time indicated in its detail
information corresponds to the time when the abnormal image occurred.
• 23BA-23BE: Error codes for the subscanning mechanism
• 2500's:
If an error is found, take remedial action as appropriate with reference to the probable
cause and troubleshooting procedures described in the error code list.
"6.1 Displaying Error Log: 1. LIST" in Maintenance Utility
SUPPLEMENTARY NOTE
The detail information format of the error code contains chronological information, from the
time when image reading started to the time when an error occurred. Reference should be
made as needed to this detail information format.
NOTE
An error for the scanner subscanning mechanism that does not accompany any
abnormal image need not be analyzed or troubleshot, and thus be ignored.
009-051-04
05.31.99 FM2460
3
4
5
6
Failure in image
processing system
conveyance system
Troubleshoot the
Troubleshoot the
controller.
conveyance system.
"3.2 Image
Troubleshooting for
Troubleshooting"
Controller"
Error codes for the scanner mechanism
CR-IR341 Service Manual
START
Failure in image
Failure in
output or image
display system
Troubleshoot the
image output
"3.3 Other Image
system.
END
MT - 178
Failure in IP and/or
X-ray exposure
system
Troubleshoot the IP
and/or X-ray
exposure system.
FR1H1519.EPS
MT - 178