FujiFilm CR-IR341 Service Manual page 536

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
Replacement
(1) Remove the cassette set unit.
"4.2 Removing and Reinstalling Cassette Set Unit"
(2) Disconnect the connector for the solenoid to be replaced from the side plate.
(3) Remove the three solenoid retaining screws to take out the solenoid.
Cassette set unit
SOLA1
SOLA2
SOLA3
SOLA4
(4) Install new solenoid.
(5) Connect the connector to the side plate.
(6) Adjust the solenoid replaced.
" Adjustment"
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
(3)
SOLA1, 2, 3, 4
Sems M3x6, x3
(2)
Connector
MC - 45
FR1H1210.EPS
MC - 45

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents