FujiFilm CR-IR341 Service Manual page 635

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
Reinstallation
NOTE
To install the scanning optics unit, loosely tighten the four screws, and position the scanning
optics unit by aligning it against the three reference pins located on the top of the subscanning
unit.
(1) Reverse the removal steps to reinstall the scanning optics unit.
Subscanning unit
(2) Take the steps described in "Check after Scanning Optics Unit Replacement."
Checks after Scanning Optics Unit Replacement
(1) Reinstall all the covers.
(2) Power ON the machine.
Make sure that the machine starts up normally.
(3) Adjust the read start position and read width.
"9.1 Adjusting Read Start Position and Read Width"
(4) Perform shading/sensitivity correction.
"9.2 Shading/Sensitivity Correction"
(5) Generate image output, and verify that there is nothing abnormal with scanner control.
Uniformly expose IPs of each size at about 1 mR. Then, check the film image recorded
under SENSITIVITY from the Test Menu for scanner control failures, such as format
misalignment and shading, which may cause diagnostic problems.
009-051-05
11.20.99 FM2638
3
4
5
6
Scanning optics unit
BR M4x16, x4
CR-IR341 Service Manual
7
8
9
10
Reference pins
FR1H1124.EPS
MC - 144
MC - 144