FujiFilm CR-IR341 Service Manual page 195

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
(13) Detailed Sequence of Search for Before-Reading Conveyance Remaining IP
Fig. 5.3.13 Search for Before-Reading Conveyance Remaining IP
Preprocessing: Fig. 5.3.1
START
Side-positioning conveyance reverse direction
DRIVE (MD3/PD72)
N
Y
Side-positioning conveyance direction stop
DRIVE (MD3/OFF) (soft)
DELAY TD74
Side-positioning conveyance direction
DRIVE (MD31/PD71)
Initialization
N
Y
side-positioning
conveyance
Side-positioning conveyance
direction stop
DRIVE (MD31/OFF)
(soft)
Before-reading standby IP found
END
Table 5.3.13 Operation Parameters
Symbol
009-051-04
05.31.99 FM2460
3
4
5
6
Is SC1 closed within TD73?
Is SD3 closed within TD75?
Description
PD71
Initialization conveyance (DOWN direction)
PD72
Initialization conveyance (UP direction)
TD73
Side-positioning conveyance remaining IP search timeout
TD74
Wait for side-positioning switchback
TD75
Initialization side-positioning
CR-IR341 Service Manual
7
8
9
10
Side-positioning conveyance
direction stop
DRIVE (MD31/OFF)
(soft)
Level 0 error
Side-positioning conveyance
03B7
remaining IP ejection error
+ infinite p
- infinite p
5.8 sec
0.5 sec
5.8 sec
TR1H0070.EPS
Side-positioning conveyance
direction stop
DRIVE (MD3/OFF) (soft)
Is SD3 closed?
Y
N
No IP
Value
MD - 132
FR1H0417.EPS
MD - 132

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr-ir341pFcr 5000Fcr 5000plus

Table of Contents