FujiFilm CR-IR341 Service Manual page 410

Fuji computed radiography
Table of Contents

Advertisement

1
2
C
ONTENTS
13A3
Preliminary steps:
1. Power OFF the machine and remove the covers.
2. Turn OFF the high-voltage switch (S1) of the SCN08A board.
3. Check for any foreign matter (lead character marking chips, etc.) inside the machine, and if any, remove it all.
Power OFF
Are the hold pin holes the cassettes
normal?
Y
Power OFF
Is the operation of the relevant shelf
solenoid arm and hold pin normal?
Lower the solenoid arm by hand to
check that it strokes properly.
Y
Power ON
Are SOLA1-4 (relevant shelf) turned
ON and the hold pin released during
initialization self-diagnostics?
Y
Power ON
Are SA3, 7, 11, and 15 (relevant
shelf) normal?
Refer to Chapter 6 Section 6.2.2 to
perform sensor checks.
Y
Restart the machine
and check its status.
009-051-04
05.31.99 FM2460
3
4
5
6
N
Replace the cassette
N
Adjust/replace the solenoid portion.
Power OFF
Is the connector connection
N
between SOLA1-4 (relevant
shelf) and DRV08A board
normal?
Replace the following
parts in order named.
Power OFF
N
Remove dust or soil from the
sensors. Power ON again to
check the sensors again.
Are SA3, 7, 11, and 15 normal?
Y
CR-IR341 Service Manual
N
Y
SOLA1—4 (relevant shelf)
DRV08A board
N
Power OFF
Are the connector
N
connections from
SA3, 7, 11, and 15
to SNS08A board
normal?
Y
Replace the following
parts in order named.
SA3, 7, 11, 15 (relevant shelf)
SNS08A board
Check the error-related
system block diagram and
connect the connectors
as appropriate.
Check the error-related
system block diagram and
connect the connectors
as appropriate.
FR1H0112.EPS
MT - 141
MT - 141