FujiFilm CR-IR341 Service Manual page 740

Fuji computed radiography
Table of Contents

Advertisement

C
11
12
ONTENTS
(7) Remove the bearing.
(8) Remove the branch path changeover guide assembly.
The arm should be removed as attached to the shaft of the branch path changeover
guide assembly.
(9) Remove the arm, bearing, and slide bearing from the branch path changeover guide
assembly.
(9)
Slide bearing
(10) Attach the arm, bearing, and slide bearing removed at step (9) to a new shock ab-
sorber-attached branch path changeover guide.
(11) Fit the spring pin into the U-shaped groove of the arm.
(12) Reinstall the branch path changeover guide assembly.
(13) Attach the bearing.
(14) Install the stopper support while the branch path changeover guide is opened (the
spring pin is pushed against the end surface of the square hole of the changeover
guide driving solenoid).
NOTES
• Make sure that the spring pin is securely pushed against the end surface of the
square hole.
• Make sure that the stopper support is in contact with the arm. Also, verify that the
spring pin or arm does not move.
009-051-04
05.31.99 FM2460
13
14
15
16
(8)
Shock absorber-attached
branch path changeover guide
Stopper support
CR-IR341 Service Manual
Erasure conveyor
(7), (9)
Bearing
BR M4x12
(9)
End surface of
the square hole
Spring pin
Arm
FR1H1821.EPS
Arm
Sems M3x10
FR1H1630.EPS
MC - 249
MC - 249