(2) Remove the guide.
Remove either of the right and left brackets that retain the guide, and then remove the
guide.
(3) Replace the two suction cups and the two packings.
NOTES
• Do not rotate the suction cup retaining screws while holding the suction cup rubber
portions. Rotate the suction cup retaining screws with a thin-blade screwdriver while
holding the pipe joint with a wrench or like tool.
• Exercise care to avoid hose disconnection.
Packing
(4) Put the IP removal arm back to the rear side.
009-051-04
05.31.99 FM2460
3
4
5
6
IP removal
arm
Hose
(3)
Packing
Suction cup
CR-IR341 Service Manual
7
8
9
10
Guide
(2)
Bracket
Sems M3x6
Suction face view
(3)
Pipe joint
MC - 72
Retaining
screw
FR1H1031.EPS
MC - 72