1
2
ONTENTS
Remedy 7
Jam location:
Cause:
IP status:
Recovery procedure
Gain access from the front or right-hand side of the machine.
a
(1) Rotate the rollers in the after-reading conveyor and erasure conveyor by hand to
position the entire IP in the after-reading conveyor. The alternative is to manually
release the after-reading conveyor roller grip and then rotate the erasure conveyor
rollers by hand to position the entire IP in the after-reading conveyor.
(2) Remove the right-hand side cover.
(3) Pull out the scanner unit and take out the IP.
If you cannot directly access the IP in the subscanning unit, first remove the light-
collecting guide and then the IP.
b
(1) Manually release the after-reading conveyor roller grip.
(2) Remove the erasure conveyor front guide and then manually convey the IP until it is
entirely positioned in the after-reading conveyor.
Remedy 8
Jam location:
Cause:
IP status:
Recovery procedure
Gain access from the front or right-hand side of the machine.
The same recovery procedure is to be performed no matter whether the IP status is a
or b.
(1) Remove the front guide from the erasure conveyor.
(2) Remove the IP by pulling it forward until it comes of the machine.
009-051-04
05.31.99 FM2460
3
4
5
6
After-reading conveyor or erasure conveyor
Conveyance failure
A certain portion of the jammed IP is located in the after-reading con-
veyor with the remaining portion placed in the erasure conveyor. There-
fore, these two units cannot be pulled out.
a) The IP cannot be conveyed forward but can be conveyed backward.
b) The IP cannot be conveyed forward or backward.
Erasure conveyor
Conveyance failure
The IP is stopped at a position immediately before the erasure conveyor
switchback point.
a) The IP cannot be conveyed forward but can be conveyed backward.
b) The IP cannot be conveyed forward or backward.
CR-IR341 Service Manual
MT - 173
MT - 173