FujiFilm CR-IR341 Service Manual page 612

Fuji computed radiography
Table of Contents

Advertisement

C
1
2
ONTENTS
8.8
Replacing Barcode Reader (BCR)
Replacement
(1) Remove the side-positioning conveyor.
"8.1 Removing and Reinstalling Side-Positioning Conveyor"
(2) Cut the four cable ties.
(3) Disconnect BCRCN2.
(4) Remove the barcode reader.
(5) Remove the four screws and take out the bracket.
Side-positioning conveyor
(6) Reverse the removal steps to reinstall a new barcode reader.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
(3)
BCRCN2
MC - 121
(5)
Barcode reader
Screw x4
(4)
Bracket
Sems M3x6, x4
FR1H1108*.EPS
MC - 121

Advertisement