10.12 Checking and Replacing Kapton
CAUTIONS
• Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
• Wear gloves during checks and replacement, because the Kapton
damaged.
Checks
(1) Remove the following covers.
Lower front cover, lower left-hand side cover
"3. Removing and Reinstalling Covers"
(2) Remove the transparent cover located on the left-hand side.
Using a blower, remove dust deposited on the transparent cover.
(3) Check the Kapton
If any soil is found, clean it off with an ethanol-moistened cloth.
If any flaw is found, replace the flawed part.
(4) Reinstall the left-hand side transparent cover.
(5) Power ON the machine.
(6) Check MZ1.
Make sure that the motor rotates normally.
"10.8 Checking and Replacing Subscanning Motor (MZ1)"
(7) Check image performance.
"3.2 Checking Image" in Preventive Maintenance Volume
009-051-04
05.31.99 FM2460
3
4
5
6
®
belt, tensioner, and flywheel for any soil or flaw.
Tensioner
Kapton
Pulley
Flywheel
(2)
Left-hand side
transparent cover
BR M4x8, x2
CR-IR341 Service Manual
7
8
9
10
®
Belt/Tensioner/Flywheel
¤
belt
Subscanning
unit
FR1H1814.EPS
MC - 195
®
belt may be soiled or
MC - 195