Adjustment
CAUTION
Before removing the covers, be sure to turn OFF the high-voltage switch. If the machine is
powered ON with any of the covers removed, the photomultiplier will be damaged.
(1) Remove the following covers.
Lower front cover, upper right-hand side cover, lower right-hand side cover, upper left-
hand side cover, upper front cover, stack table, multi-stage cassette loading cover
"3. Removing and Reinstalling Covers"
(2) With the cassette set unit pulled out, connect the connectors (CNA1, 2) of the cassette
set unit.
"4.2 Removing and Reinstalling Cassette Set Unit"
Cassette set unit
(3) Power ON the machine.
The initialization sequence is started, and, after 30 to 60 seconds have passed, a
software version number (Axx) appears on the operation panel.
(4) Immediately after the software version number (Axx) appears, touch the upper left and
upper right corners of the operation panel in sequence.
M-Utility is activated.
If "STEP 95" is indicated, any touch on the operation panel is rejected. In that case,
reset the machine and wait until the software version number (Axx) appears on the
operation panel.
(5) Select "5. MECHANICAL UTILITY" and "1. INITIALIZE" in sequence.
(6) Select "4. SENSOR" and "2. MONITOR".
A list of sensor Nos. is displayed.
009-051-04
05.31.99 FM2460
3
4
5
6
CR-IR341 Service Manual
7
8
9
10
MC - 36
CNA1
CNA2
FR1H1058.EPS
MC - 36